13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.7
Adjustment of the ceph secondary diaphragm (QuickShot)
13.2.1.7.1
Automatic adjustment: Ceph secondary diaphragm (QuickShot)
(standard)
Phantom for Ceph exposure
✔
The pan needle phantom is
not
in the pan bite block holder on the
unit.
✔
The ceph test phantom does
not
connect to the cover of the sensor
unit.
1. In the structure tree, under
"Ceph"
, click on the
"Adjustment of Ceph
secondary diaphragm (Quickshot)"
element (S010.8).
The
"Adjustment of Ceph secondary diaphragm (Quickshot)"
menu is displayed in the action area.
2. Establish [ → 209] receptivity.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.