64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
261
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
3. Adjust the ear plugs in the vertical direction by turning the knurled nut
(F). The direction of rotation for left-arm (G) and right-arm (H)
versions is shown in the drawing below.
4. Tighten the screw (E) firmly.
Following adjustment
1. Swing the ear plug holders back into the beam direction (black
adjusting cap on the outside).
Summary of Contents for ORTHOPHOS SL 2D