13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.2.2
Diaphragm adjustment
✔
The 3D calibration block
is not
in the bite block holder on the unit.
1. In the structure tree, under
"3D"
, click on the
"Diaphragm
adjustment"
element (S030.10).
2. Establish readiness for exposure [ → 209].
3. Create an exposure (85 kV / 0.34 mAs) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the values can be applied.
4. Repeat the procedure starting at Step 2 until adjustment is
successful.
5. When adjustment is OK, save the values. [ → 209]
A message is displayed in the action window to indicate that the
values have been successfully saved.
The status indicator that prefixes the element
"Diaphragm
adjustment"
is checked.
6. Continue with the next stage of the calibration procedure. [ → 265]