13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
8. Establish [ → 209] receptivity.
9. Take an exposure (85 kV/22 mA s) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
of the
"5x5"
/
"Premolar left"
element in the structure tree
responds accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
10. Repeat the procedure starting at Step 8 until calibration is successful.
11. Repeat calibration for all centers of rotation of volume
"5x5"
and for
all other volumes and their centers of rotation.
12. Remove the 3D calibration block from the pan bite block holder on the
unit.