64 95 142 D3632
276
D3632
.
031.01.02
.
02 03.2016
14 Final work
Sirona Dental Systems GmbH
14.1 Save sensor data
Installation Manual ORTHOPHOS SL
14
Final work
14.1
Save sensor data
Backing up sensor data1
You can use the
"Backup sensor data"
menu to load the sensor data
(factory calibration of the sensor) from the DX83 board and save it as a
Zip file.
NOTE: Since the sensor data is saved on the DX83 board, this is required
e.g. before exchanging the DX83 board. The saved sensor data must
then be resaved on the new DX83 board following replacement of the
board. This is possible via the
"Restore sensor data"
menu (see the
Service Manual).
Open the
"Backup / Restore sensor data"
Open the backup / restore data menu
✔
The unit and computer are turned on.
✔
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
NOTICE! The IP address of your unit can be found on the Info screen.
3. Confirm your input with the "Enter" key.
A security prompt appears.
4. Click on the link
"Continue to this Website"
.
The Sirona Web service is loaded.
The sub menus of the
"User"
menu are displayed in the structure
tree.