7 Installation: Remote control
Sirona Dental Systems GmbH
7.5 Connecting the X-ray warning lamp
Installation Manual ORTHOPHOS SL
7.5
Connecting the X-ray warning lamp
You can activate an X-ray warning lamp via the remote control.
1. Guide the cable (A) for connecting the X-ray warning lamp into the
chassis (concealed or surface installation).
2. Attach the cable to the still exposed strain relief.
3. Connect the current-carrying cables and the ground wire to terminal
X105 (board DX42) as shown in the connection diagram.
4. Secure the current-carrying cables with a cable tie (B) to prevent
them from slipping out.
5. Lay the cable (A) parallel to control cable L117 and close the housing.
NOTICE
A maximum load of 50W is permissible. No additional circuit may be
connected.
Use a 3-wire cable (1.5mm²) to connect the X-ray warning lamp.
A
C
X-ray warning lamp
X105
DX42
X108
X105
K100
DX42
C
L117
B
A