Sirona Dental Systems GmbH
10 Installation: Ceph arm
Installation Manual ORTHOPHOS SL
10.3 Installation
båÖäáëÜ
6. Switch on the unit (G).
During the unit startup phase, the start screen appears on the
control panel.
The unit's self-adjustment routine begins.
7. Acknowledge the error messages that appear using the R key (H) on
the control panel.
The main menu and help message H301 appear. You are
requested to move the unit to its starting position.
8. Press the R key (H) on the control panel.
The unit moves to its starting position.
9. Use the arrow keys (I) on the control panel to lower the unit until the
bearing bolts on the unit are at the same height as the drill holes on
the ceph support arm.
NOTICE! Check the room height before you raise the unit. If the room
height is less than 2.27 m (89 3/8") or 2.30 m (90 1/2") for installation
with the floor stand, you must limit the maximum travel height (S018:
Installing the ceph arm
1. Attach the support arm to the four bearing bolts.
NOTICE! The connecting cables must not be crushed. Make sure
that the connecting cables lie in the groove of the support arm and
are not crushed.
2. Screw in the conical nuts (J) and tighten them firmly.
3. Remove the styrofoam packaging from the ceph.
4. Follow the instructions in the section entitled “Installing the secondary
G
I
H
IMPORTANT
The support arm may get tilted while you are tightening the conical nuts.
You can avoid this by lifting and moving the support arm gently while you
are tightening the nuts.
4x
J