11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.6.4
Possible results of self-test and troubleshooting measures
"Success"
The self-test was successfully executed. No errors were detected in the
image data path.
"Timeout"
The self-test could not be completed within 20 seconds.
"Unknown"
The self-test was canceled with an error that could not be localized.
"Prep"
An error occurred during the activation of the exposure for the self-test. If
no further error messages (see title bar of Easypad) occur, please
implement the following tests and measures:
● Check connecting cable L83 between board DX1 and the sensor unit.
● Replace board DX11
● Replace sensor unit
If error messages are present, please implement the measures according
to the displayed error message.
"Transfer"
Minimal image data was received by the sensor unit during the self test.
If no further error messages (see title bar of Easypad) occur, please
implement the following tests and measures:
● Check connecting cable L83 between board DX1 and the sensor unit.
● Replace board DX11
● Replace sensor unit
If error messages are present, please implement the measures according
to the displayed error message.
"Command"
A timeout occurred at the sensor unit during a command on the DX11
exposure. If no further error messages (see title bar of Easypad) occur,
please implement the following tests and measures:
● Restart unit perform self test once again.
If the error re-occurs:
● Check connecting cable L83 between board DX1 and the sensor unit.
● Replace board DX11
● Replace sensor unit
If error messages are present, please implement the measures according
to the displayed error message.