64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
177
Sirona Dental Systems GmbH
12 Startup, measurements and tests (for USA/Canada only)
Installation Manual ORTHOPHOS SL
12.1 Safety
båÖäáëÜ
12
Startup, measurements and tests (for USA/
Canada only)
12.1
Safety
NOTICE
Qualifications of service personnel
Installation and startup may be carried out only by personnel specifically
authorized by Sirona.
WARNING
Perilous shock hazard!
You must switch off the unit and then wait at least 1 minute, for
measurements at the tube assembly at least 4 minutes, before you start
to connect test cables or remove a cover!
For measurements in the area of the power supply terminal, the unit
must be disconnected from the junction box of the building installation
before you start to connect test cables!
DANGER
X-rays
When performing the following tests, be sure to observe the radiation
protection regulations applicable in your country (see Operating
Instructions).
DANGER
X-rays
"Radiation" is signaled by the message "X-RAY active!", a beep, and an
X-ray LED.
NOTICE
Risk of damage to boards
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
IMPORTANT
It is essential that you also observe the notes about the operation of the
unit in the Operating Instructions.
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......