13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
13.1.6.3
Geometry phantom for volume calibration
* For greater clarity, the unit is shown in the left-hand image without
sensor.
For
"Sensor adjustment"
3D adjustment and
"Geometry calibration"
3D
calibration, you must insert the geometry phantom (A) into the pan bite
block holder (B) on the unit.
For
"Diaphragm adjustment"
3D adjustment, as well as for
"Sensor
calibration (DCS)"
and
"Diaphragm calibration"
3D calibrations, you
must take the geometry phantom back out of the bite block holder.
NOTICE
Risk of damage to unit
Make absolutely sure that the geometry phantom is inserted in the bite
block holder of the unit in such a way that the body of the phantom and
the screw (C) point
away from the unit in the direction of the patient
(see
figure). Otherwise, this can result in a collision between the sensor unit
and the geometry phantom and thereby damage the unit.
*