13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.2.6
Dosimetry
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
1. In the structure tree, under
"3D"
, click on the
"Dose measurement"
element (S011.21).
2. Attach the Mult-O-Meter sensor in the middle of the sensor unit (flat
panel detector side).
3. Select a program from the 3D program group on the control panel.
The unit moves to the 3D exposure position.
4. Establish [ → 209] receptivity.
5. Create an exposure (85kV / 35 mAs).
The status indicator that prefixes the
"Dose measurement"
element is checked.
6. Read the dose shown on the Mult-O-Meter.
The dose value must be between 1.2 and 2.3 mGy.
7. Note down the dose for later recording.
8. Remove the Mult-O-Meter from the sensor unit again.
9. Click the
"Exit"
button to close the calibration menu.