64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
77
Sirona Dental Systems GmbH
7 Installation: Remote control
Installation Manual ORTHOPHOS SL
7.4 Connecting the control cables (REMOTE)
båÖäáëÜ
7.4.2
Installation version 2: With spiral cable
1. Use an awl to puncture the membrane keyboard at the prepared
points (A) from the rear.
2. Using the two screws (C), fasten the release button holder (B) to the
keyboard.
3. Plug the connector of the coiled cable into socket X101 on board
DX42 and screw the connector down tight.
4. Hook the coiled cable onto the strain relief (D) of the assembled
chassis.
5. Connect control cable L117 (see section “Installation version 1:
Without coiled cable [ → 76]”).
CAUTION
Operation of the remote control via the membrane keyboard is
prohibited when installing the remote control with release button.
A
C
B
X101
D
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......