båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 49
Tab 2
2.5 List of error messages
2.5
*)
TTP = Time Trigger Protocol
E7
71
21
No CAN bus connection.
DX11 does not start.
Note:
Occurs after power-on in the start
screen.
z
Start the detail query via Sixabcon
If DX11 responds...
z
Check signal path to DX71,
repair or replace cable/connector if necessary
z
Replace DX1
If DX11 does not respond...
z
Replace DX11
E3
71
30
Height adjustment keys actuated dur-
ing power-on
z
Unit restart:
Switch the unit OFF. Wait for 30 sec.
Switch unit ON, making sure that the Multipad is
not actuated during boot-up.
If the error occurs again...
z
Replace board DX71
or
Multipad
E3
71
31
Forehead support key pressed during
power-on
E3
71
32
Temple support keys actuated during
power-on
E3
71
33
Light localizer key actuated during
power-on
E3
71
34
Test key actuated during power-on
E3
71
35
Return key actuated during power-on
E3
71
36
Service key actuated during power-on
E3
71
37
Memory key actuated during power-
on
E3
71
38
Program selection key actuated during
power-on
E3
71
39
Radiation time key actuated during
power-on
E3
71
40
kV/mA key actuated during power-on
E3
71
41
Patient symbol key actuated during
power-on
Error code
Description
Actions required
see page
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...