Troubleshooting
DANGER
Shock hazard! It is essential that you switch the X-ray unit OFF before
replacing any components!
CAUTION
– Switch the X-ray unit OFF before connecting a measuring instrument.
– Select the correct current/voltage type and adjust the measuring range to
match the expected readings.
– Perform continuity tests only on units which are switched off.
– Observe the prescribed cool-down intervals if several exposures must be
taken to check a measurement.
CAUTION
Please observe the usual precautionary measures for han-
dling printed circuit boards (ESD).
Touch a ground point to discharge static electricity before
touching any boards.
CAUTION
CAN bus cable:
When unplugging CAN bus cables, it is essential to unplug
the power supply as well.