3.8 Error analysis of X-RAY control signal path: up to unit hardware revision AG
3.8
Error analysis of X-RAY control signal path:
up to unit hardware revision AG
(with board DX41)
3.8.1
Error and help messages
with remote control
installed
E3 42 31 + E3 13 40 + E3 41 20
occur in combination after the unit
is switched ON with the door
contact closed.
E3 42 31
occurs individually after the unit is
switched ON.
yes
no
Was the release but-
ton actuated during
switch ON?
Switch the unit OFF
and then ON again.
Make sure that the
release button is not
actuated during
switch-ON.
Do the error messages
recur?
yes
Short circuit in the coiled cable of the release
button or on the membrane keyboard of the
remote control:
• Replace release button
• Replace remote control
Hardware fault on display board DX42 or short circuit in coiled
cable of release button or on membrane keyboard of the remote
control:
• Replace release button (unit hardware version
<
BA)
• Replace remote control (see section 6.20)
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...