59 38 399 D3352
3 – 24
D3352.076.01.13.02
07.2008
3.9 Error analysis of X-RAY control signal path: unit HW version BA or higher
3.9.3
Error messages
with and without installed
remote control
Error code
Description
Actions required
see page
E3
13
40
Short circuit in signal path between
release button A2 and board DX11
during switch-ON.
z
Replace cable L117 or L108
z
Replace board DX1
z
Replace board DX11
E6
13
41
Release signal missing on board
DX11 at start of exposure.
z
Replace cable L117 or L108
z
Replace board DX1
z
Replace board DX11
E3
13
42
Short circuit in signal path between
release button A2 and board DX11
during unit operation.
z
Replace cable L117 or L108
z
Replace board DX1
z
Replace board DX11