3.10 Check data paths/Generate test images
3.
10
7.
Check the images according to the following criteria:
– Tile structure of the existing segments
NOTE
i
To facilitate checking, you may adjust the image contrast and brightness in
SIDEXIS.
– Linear grayscale gradient
– Intensity steps clearly recognizable
– No image artifacts detectable
PAN test image
CEPH test image
TSA test image