07.2008
4 – 17
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
Making the unit ready for exposure
7.
Confirm the sensor adjustment
and make SIDEXIS XG ready for exposure:
Click
OK
SIDEXIS screen
The exposure dialog box showing the exposure status
appears in SIDEXIS.
XG
Plus
:
Display on the Easypad
The Easypad touchscreen shows service routine
S010.1
as well as the related exposure parameters.
XG 5 / 3 / 3 PPE:
Display on the Multipad
The single-line display of the Multipad shows the exposure parameters
of service routine
S010.1
.
1
2
3
+
-
60 kV/3 mA
0.60 s
S10
1
1
2
3
60/3____0.60____