4.3 Adjusting the panoramic X-ray unit
4.
3
NOTE
i
If you do not reach your goal via automatic adjustment, repeat the adjustment
procedure with manually determined adjustment values (see page 4-20).
Saving the values
10.
If the image is identical to the ideal image
(A)
, save the values:
Click
SAVE
VALUES
NOTE
i
In SIDEXIS versions higher than V01.50. the values for
S1 - S3
in the
PAN
-
SENSOR
ADJUSTMENT
submenu are
setequal to zero
with a correct adjust-
ment, i.e. if they are within the permissible tolerance. For versions < V01.50,
the values in the text boxes may deviate slightly from zero even with a correct
adjustment.
z
Remove the needle phantom from the bite block holder of the unit.
ORTHOPHOS XG 3 / 3 PPE:
z
Reattach the fixed diaphragm (see page 4-10).
z
Go on to the next adjustment step.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...