59 38 399 D3352
4 – 22
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
4.3.4
Adjusting the (motorized) pan diaphragm
(XG
Plus
/ XG 5)
Menu: Diaphragm/system adjustment
‡
Pan - Diaphragm
Preparations
z
Remove the needle phantom from the bite block holder of the panoramic
X-ray unit (see page 4-9).
NOTE
i
If a message window indicating that the diaphragm is tilted appears during the
pan diaphragm adjustment, this means that the primary diaphragm is me-
chanically maladjusted in the vertical axis. In this case, contact the Sirona
Customer Service Center:
Opening
PAN
-
DIAPHRAGM
1.
Go to the
PAN
-
DIAPHRAGM
submenu.
NOTE
i
The menu provides a precision adjustment and a coarse adjustment
(precision adjustment is preset). Perform a precision adjustment first. In most
cases, previous coarse adjustment is not necessary.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...