4 – 41
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
Evaluating the X-ray image
4.
Evaluate the X-ray image:
– The superimposed filter must cover one half of the diaphragm (A).
NOTE
i
If these criteria are not fulfilled
(B)
, the pan filter must be adjusted.
Making the unit ready for exposure
NOTE
i
The default value for Fx was automatically determined by SIDEXIS based on
the exposure and entered in the text box of the menu.
For manual adjustment, the value displayed at this position in the text box of
the menu can be overwritten (see page 4-44).
First continue with the automatic adjustment. Manual determination of the ad-
justment value is required only if you fail to reach your goal via automatic ad-
justment (see page 4-44).
5.
Make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
Adjustment: ok
B
Adjustment: not ok
A
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...