59 38 399 D3352
4 – 54
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
4.3.8
Checking and adjusting the touchscreen
(XG
Plus
only)
Switch the unit ON
1.
Switch the unit ON.
2.
When the start screen appears (see above), press the
light localizer
,
T
and
R
keys immediately and simultaneously.
NOTE
i
These keys must be pressed while the unit is booting and the start screen is
displayed.
The
first
adjustment screen appears on the display.
Performing a calibration
3.
Calibrate the touchscreen:
z
Touch the
upper left
corner.
The
second
adjustment screen appears on the display.
z
Touch the
lower right
corner.
The
third
adjustment screen appears on the display.
see next page