59 38 399 D3352
4 – 78
D3352.076.01.13.02
07.2008
4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed) Tabs 4
Saving the values
13.
If the image is identical to the ideal image
(A)
, save the values:
Click
SAVE
VALUES
NOTE
i
In SIDEXIS versions V01.50 and higher, the values for
S1
,
S2
and
S3
in the
CEPH
-
FIXED
POINT
OF
ROTATION
submenu are set
equal to zero
with a cor-
rect adjustment, i.e. if they are within the permissible tolerance. For versions
< V01.50, the values in the text boxes may deviate slightly from zero even with
a correct adjustment.
z
Go on to the next adjustment step.