4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed) Tabs 4
Evaluating the X-ray image
7.
Evaluate the X-ray image:
– The exposed diaphragm area must lie centered and straight in
the image field as well as inside the superimposed auxiliary lines
(A).
– A white border surrounding the image on all sides must be visible.
The maximum density must lie in the center of the diaphragm area
(A).
NOTE
i
If this criterion is not yet fulfilled, repeat the adjustment procedure starting with
step 5.
If you do not reach your goal via automatic adjustment, repeat the adjustment
procedure with a manually determined adjustment value (see page 4-98).
Adjustment: ok
A