Tab 4 4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed)
4.
4
Removing the cover
3.
Unscrew and remove the cover from the cephalometer.
Opening the SIDEXIS Service menu
4.
Open the SIDEXIS Service menu: To do this, proceed as follows:
Opening the constancy test
z
Select the constancy test in SIDEXIS XG:
Select
EXTRAS
‡
CONSTANCY
TEST
The typical SIDEXIS user interface is started.
Constancy test is already preset.
Starting the exposure mode
z
Start the exposure mode:
Click
XCXP
The dialog box for selecting the X-ray device appears on the screen.
Selecting/confirming the X-ray device
z
Select/confirm the desired X-ray device:
Select e.g.
KAMMER
4
and click
OK
The dialog box for selecting the test type appears on the screen.