4.6 Resetting the adjustment (XG
Plus
/ 5 / 3 / 3 PPE)
4.
6
4.6
Resetting the adjustment
(XG
Plus
/ 5 / 3 / 3 PPE)
NOTE
i
Important: Make sure to note down the values displayed in the text boxes
before modifying them. This will enable you to reset the adjustment val-
ues to the factory settings if necessary.
Contact the Sirona Customer Service Center for more information (or to
enable the menu):
Phone: 0 62 51/16 - 16 16
The
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu offers you the possibility of
resetting or manually modifying any or all of the pan or ceph adjustment
settings you have made
in very exceptional cases
.
Resetting the pan adjustment
z
To reset the pan settings, open the
PAN
-
RESET
ADJUSTMENT
menu.
Ceph - Reset adjustment
(not for XG 3/3 PPE)
z
To reset the ceph settings, open the
CEPH
-
RESET
ADJUSTMENT
menu.
NOTE
i
If the adjustment values have been reset, the unit must be readjusted.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...