07.2008
5 – 21
Tab 5
5.4 Service routine S001
5.
4
S001: Radiation without rotary movement, fixed radiation time
5.4
Service routine S001
Radiation without rotary movement, fixed radiation time
WARNING
Be sure to observe the radiation protection regulations applicable in
your country.
z
Select the Service menu (see page 5-10).
Selecting service routine S001
1.
Use the
arrow keys
in selection field 1 to select
service routine
S001
and confirm this selection with the
Service key
.
Confirming the security access
2.
Confirm the security access by once again selecting the number of the
main routine (1) with the arrow keys of selection field 2 and then pressing
the
Service key
.
Function S001:
Test of X-ray beam and kV/mA levels
Radiation with selected kV/mA level
Radiation time cannot be selected, max. 14.0 sec.
Primary diaphragm remains in last position selected
Security access
1
2
3
+
-
S001
n
(
1
2
3
S001____________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
(
o
(
o
1
2
3
-
+
1
n
(
1
2
3
___________1____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
(
o
(
o
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...