5 – 23
Tab 5
5.5 Service routine S002
5.
5
S002: Radiation without rotary movement, maximum radiation time can be selected
5.5
Service routine S002
Radiation without rotary movement, maximum radiation time
can be selected
WARNING
Be sure to observe the radiation protection regulations applicable in
your country.
z
Select the Service menu (see page 5-10).
Selecting service routine S002
1.
Select service routine
S002
using the
arrow keys
of selection field 1.
Function S002
X-ray beam test
Radiation with selected kV/mA level
Maximum radiation time selectable
Radiation from any position
Test step 1
Primary diaphragm remains in last position selected
Security access
Test step 3
Primary diaphragm opened fully
Security access
Test step 4
Filter element for 0.8 Cu step filter
The primary diaphragm is set to a widened opening symmetrical to the
PAN setting so that the step filter still remains covered.
Security access
1
2
3
+
-
S002
1
n
(
1
2
3
S002_______1____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
+
-
(
n
(
n
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...