59 38 399 D3352
5 – 28
D3352.076.01.13.02
07.2008
5.6 Service routine S005
Tabs 5
S005: General X-ray tube assembly service
5.6
Service routine S005
General X-ray tube assembly service
WARNING
Be sure to observe the radiation protection regulations applicable in
your country.
z
Select the Service menu (see page 5-10).
Selecting service routine S005
1.
Use the
arrow keys
to select service routine
S005
.
Function S005
General X-ray tube assembly service
Test step 2
Automatic compensation of preheating
Security access
Test step 4
Fan test
Test step 5
Temperature sensor test
Test step 6
Switch off cool-down interval of the single tank
Security access
Test step 7
Configuring continuous operation of the tube assembly fan
Security access
1
2
3
+
-
S005
2
n
(
1
2
3
S005_______2____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
+
-
(
n
(
n
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...