5 – 35
Tab 5
5.6 Service routine S005
5.
6
S005.6: Switch off cool-down interval of the single tank
The status of the pulse/pause treatment is displayed in the selection
field 1:
01 = Pulse/pause monitoring enabled
00 = Pulse/pause monitoring disabled
Disabling the pulse/pause monitoring
CAUTION
The execution of this service routine is irreversible. The system is
operated beyond its specifications afterwards. This may result in an
X-ray tube assembly failure. As a result, the system warranty expires.
SIRONA is able to prove this system setting after a system failure.
3.
To distable the pulse/pause monitoring, select code 00 in selection field 1
with the arrow keys.
The
Memory key
(XG
Plus
) or the LED above the
Memory key
(XG 5 / 3 / 3 PPE)
lights up.
1
2
3
-
+
01
6
n
(
1
2
3
______01________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
R
1
2
3
-
+
00
6
n
(
1
2
3
______00________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
R
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...