båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 61
Tab 5
5.10 Service routine S012
5.10
S012.2: Inquiry of the CAN status register of the modules
Selecting the register
3.
Press the
Test key
.
XG
Plus
:
Easypad touchscreen
The CAN status registers of the currently selected module,
e.g. "CAN-State DX6" are displayed.
NOTE
i
If values deviate from zero, this indicates a problematic CAN bus connection.
XG 5 / 3 / 3 PPE
: Multipad
The first CAN status registers of the currently selected module,
e.g. "CAN-State DX6" are displayed.
Select the desired CAN status register using the
arrow keys
in selection field
1 and confirm your selection with the
Memory key
.
The value of the selected CAN status register is displayed.
You can return to the CAN status register selection menu using the
up arrow key
or the
Service key
.
T
T
T
T
(
n
)
n
1
2
3
+
-
DX6
n
(
1
2
3
DX6____________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
(
n
(
n
)
65
CAN-State DX6
CAN-Error Comp.
:
0
Overrun Error
:
0
Receiver Warn.
:
0
Receiver Error
:
0
Transmit Warn.
:
0
Transmit Error
:
0
Bus off
:
0
n
(
1
2
3
C A N
_
E R R O R
_
CO MP
.__
(
n
(
o
n
(
1
2
3
0_____________
(
n
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...