båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 103
Tab 5
5.13 Service routine S017
5.13
S017.17: Setting the standard acceptance/constancy test, RöV FRG
5.13.14
S017: Test step 17
Setting the standard acceptance/constancy test,
RöV FRG
Selecting test step 17
1.
Select test step 17 using the
arrow keys
in selection field 2 and confirm
your selection with the
Service key
.
NOTE
i
In
XG
Plus
and
XG 5
units, the default value is already displayed correctly as
soon as service routine S017.17 is opened, and therefore only need by stored.
In
XG 3
units, the standard must be set as described in the following step.
1
2
3
+
-
S017
17
XG
Plus
:
Easypad touchscreen
+
-
(
o
S17
(
n
n
(
1
2
3
S017______17____
XG 5 / 3 / 3 PPE:
Multipad
(
n
(
o
1
2
3
+
-
02
XG
Plus
:
Easypad touchscreen
R
S17
n
(
1
2
3
______01________
XG 5 / 3 / 3 PPE:
Multipad
17
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...