59 38 399 D3352
5 – 108
D3352.076.01.13.02
07.2008
5.13 Service routine S017
Tabs 5
Configuration of PPE display warning level
5.13.17
S017: Test step 23
(XG 3 PPE only)
z
Start test step 23 in the same way as described on page 5-78 or by
pressing the
Service key
in test step 22.
NOTE
i
If you go to test step 23 from test step 22 by pressing the Service key, the
security access confirmation will be skipped.
Defining the warning level for display of
remaining exposures after each
exposure
1.
Use the
arrow keys
in selection field 1 to select the number of remaining
exposures starting with which the warning (indicating the number of expo-
sures left) will be displayed after each exposure.
2.
To save your selection, first press the
Memory key
(the R key lights
up) and then press the
R key
.
Quitting the service routine
To return to the Service menu, press the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
n
(
1
2
3
______30________
XG 3 PPE:
Multipad
R
(
n
R
n
(
1
2
3
______50________
XG 3 PPE:
Multipad
(
o
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...