5.14 Service routine S018
Tabs 5
S018.2: Set minimum travel height
Saving the minimum travel height
3.
To save the minimum travel height, first press the
Memory key
(R key (XG
Plus
) or LED above R key
(XG 5 / 3 / 3 PPE) lights up) and then the
R key
.
NOTE
i
The limitation of the minimum travel height is purely software based. The lower
limit switch is not adapted to the new minimum travel height!
Quitting the service routine
XG
Plus
:
To return to the service routine selection menu, press the
Service key
or the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
For XG 5 / 3 / 3 PPE:
Pressing the
up arrow key
above selection field 3 returns you to the ser-
vice routine selection menu.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
R
R
(
n
R
(
o
1
2
3
200mm
n
(
1
2
3
__200_
MM
________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
R
)
S18
5
n
)
o
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...