59 38 399 D3352
5 – 120
D3352.076.01.13.02
07.2008
5.15 Service routine S020 (XG
Plus
only)
Tabs 5
S020.1: Change switching points of the temple support
2.
Select the desired offset for the switch-over point using the
arrow keys
of
the corresponding selection field (see page 5-118).
Selection field 1: Lower switching point (L)
Selection field 2: Upper switching point (U)
NOTE
i
A negative offset shifts the respective switching threshold toward larger skulls;
a positive offset has the opposite effect.
After the offset value is changed, the
Memory key
lights up.
3.
To save the offset values, first press the
Memory key
(R key lights up) and then the
R key
.
Quitting the service routine
Pressing the
Service key
or the
double arrow key
returns you to the
service routine selection menu.
To quit the Service menu and return to the main menu, press the
double arrow key
.
R
1
2
3
+
-
L: -3
U: 0
XG
Plus
:
Easypad touchscreen
+
-
1
(
n
R
R
1
2
3
+
-
L: -3
U: 0
XG
Plus
:
Easypad touchscreen
+
-
S20
1
)
n
)
o
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...