5.18 Service routine S033 (not for XG 3 / 3 PPE)
Tabs 5
S033: Test of CEPH image path without SIDEXIS
5.18
Service routine S033
(not for XG 3 / 3 PPE)
Test of CEPH image path without SIDEXIS
z
Select the Service menu (see page 5-10).
Selecting service routine S033
1.
Use the
arrow keys
to select service routine
S033
.
5.18.1
S033: Test step 10
Selecting test step 10
1.
Select test step 10 in selection field 2 with the
arrow keys
and confirm
your selection by pressing the
Service key
.
After test step 10 is selected, selection field 1 displays the current selection
for the synchronized readout sequence.
Function S033
Test of CEPH image path without SIDEXIS
Test step 10
Test of the CEPH image path without SIDEXIS
1
2
3
+
-
S033
n
(
1
2
3
S033_______10___
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
(
n
(
n
1
2
3
+
-
S033
10
n
(
1
2
3
S033_______10___
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
+
-
(
n
(
o
(
o
S33
(
n
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...