5.19 Service routine S034
5.19
S034.4: Display/calibrate center position of ceph scan sensor axis
5.19.1
S034: Test step 4
z
Remove the cover from the cephalometer (see also section 1.14).
Selecting test step 4
1.
Select test step 4 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
After test step 4 is selected, the current offset value for the center position of
the ceph scan
sensor axis
is displayed in selection field 1.
1
2
3
+
-
S034
4
n
(
1
2
3
S034_______4____
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
+
-
(
n
(
o
(
o
S34
(
n
1
2
3
1500
n
(
1
2
3
____1500________
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
(
n
S34
4
R
T