59 38 399 D3352
5 – 140
D3352.076.01.13.02
07.2008
5.19 Service routine S034
Tabs 5
S034.5: Display/calibrate center position of ceph scan secondary diaphragm axis
Determining the offset values
3.
Move the secondary diaphragm to its front position by pressing the
R key
.
NOTE
i
The secondary diaphragm can always be moved back and forth between its
front and center positions with the R key.
4.
Move the secondary diaphragm to its center position by pressing the
R key
.
5.
Measure distances
L2
and
L1
with a slide gage.
6.
Calculate the offset value according to the following formula:
Ceph arm on left side:
Ceph arm on right side:
Secondary diaphragm side
Wall side
Room side
R
R
R
R
Wall side
Room side
L1
Ceph arm mounted on
left-hand side
Ceph arm mounted on
right-hand side
L2
L2
L1
L2
L1
)
2
÷
–
(
Offsetwert
=
L1
L2
)
2
÷
–
(
Offsetwert
=
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...