6.9 Replacing/adjusting the FH light localizer (Ceph)
6.9
Adjusting the Ceph light localizer (FH)
z
Switch the unit
ON
.
z
Switch the light localizers
ON
.
z
Align the Ceph (FH) laser module so that FH light beam
C
runs horizontal
between the ear plugs (tolerance: ± 1.5 mm).
z
Tighten screw
A
securely.
NOTE
i
Be careful not to turn the laser module when you tighten the screws.
z
Check the light localizer again and perform a correction if necessary.
Attaching the covers
z
Reattach the cover of the secondary diaphragm.
z
Reattach the cover to the cephalometer.
ATTENTION: What to do after replacement?
– No further action is required.