båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
6 – 33
Tab 6
6.11 Replacing the support piece (bite block holder)
6.11
6.11
Replacing the support piece
(bite block holder)
Removing the defective support piece
1.
Loosen the two screws
(A)
and remove the support piece.
Installing the support piece
z
Place the support piece on the tube bend and fasten it with the two screws
(A)
.
ATTENTION: What to do after replacement?
– Check the system adjustment via a diaphragm test exposure
PAN - Symmetry (see page 7-22).
– Perform a complete PAN unit adjustment(see chapter 4).
A
1.
Summary of Contents for ORTHOPHOS XG 3 DS