59 38 399 D3352
6 – 52
D3352.076.01.13.02
07.2008
6.19 Replacing the light barriers
Tabs 6
NOTE
i
The three light barrier cables are bundled with a flexible tube and fastened to
the basic unit next to board DX1 or on the upper wing plate with two
cable clamps.
z
Detach the cable harness (flexible tube) from the cable clamps and re-
move the flexible tube.
z
Remove the ring motor (see page 6-14).
3.
Loosen the screw
(C)
on the wing plate
(Caution: The eccentric spring
is prestressed!)
. Unhook the spring and release it carefully. Remove the
eccentric spring
D
.
4.
Detach the remaining screws from the wing plates and initially remove the
upper wing
E
.
3.
C
D
4.
E
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...