båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
6 – 57
Tab 6
6.19 Replacing the light barriers
6.19
Cabling V1_1, V1_2 and V1_3
z
Run the light barrier cables together next to the ring motor and pull the
flexible tube over them.
11.
Secure the flexible tube at the motor end with a cable tie and fasten it to
the wing plate with the cable clamp.
12.
Twist the flexible tube (only the tube, not the cables!) clockwise at the con-
nector end, secure it with a cable tie and fasten it to the basic unit next to
board DX1 with the cable clamp.
13.
CAUTION
In its installed state, the cable harness (flexible tube) must exhibit a tendency
to move upward and form a loop during travel of the rotary ring. It must dip into
the unit!
Also make sure that the cable harness is refastened to the cable clamps in its
original position.
Properly run cable
harness
11.
12.
13.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...