7.1 Checking the height adjustment
7.
1
Check whether the height adjustment
limit switches are functioning properly
z
Manually actuate both limit switches one after the other while the height
adjustment motor is running: The motor must stop.
If the limit switches are not functioning:
Æ
Check the corresponding microswitch and replace if necessary
Æ
Check cable L19, replace if necessary
Check whether an audible signal can be
heard during height adjustment
z
Run the unit up and down: An audible signal must sound.
If no audible signal sounds:
Æ
Replace board DX1 (see Page 6-59)
Actuator
Limit switch
50