7.3 Checking the sensor holder (pan and ceph)
Tabs 7
Check whether the sensor locks and
unlocks easily and is firmly seated in
the holder
z
Carefully insert the sensor upward into the holder:
The sensor must snap in place audibly.
z
Check the sensor for firm seating.
z
Press the locking button and carefully pull the sensor downward out of the
holder: The sensor must be easy to remove.
If the sensor cannot easily be inserted in or removed from the holder, or is
seated too loosely in the holder:
Æ
Check the magnets on the sensor for dirt or foreign particles and clean
them or remove any foreign particles if necessary
Æ
Replace the sensor holder (see Page 6-43)
Æ
Replace the sensor
Summary of Contents for ORTHOPHOS XG 3 DS