7.8 Checking the diaphragm
Tabs 7
Evaluating the X-ray image
4.
Evaluate the X-ray image:
– The exposed diaphragm area must lie horizontally centered in the
image field (A).
– A white border surrounding the image on all sides must be visible.
The maximum density must lie in the center of the diaphragm area
(A).
NOTE
i
If these criteria are not fulfilled
(B)
, the TSA sensor must be adjusted (see
chapter 4).
Adjustment: ok
Adjustment: not ok
A
B