7.14 Checking the device leakage current
Tabs 7
Measuring setup for testing the device leakage current
z
Check whether the power switch of the unit is turned
ON
.
z
Connect a high-resistance measuring voltage source between the short-
circuited power cable
(B)
and ground wire
A
.
z
Measure the voltage drop across
MD
.
NOTE
i
The measured value must not exceed
5 mA
.
If the leakage current is not OK, perform troubleshooting according to section
3.6.
z
Reconnect the unit to the building installation (see the Installation Instruc-
tions for the ORTHOPHOS XG).
Power cable
to the Orthophos XG
B
V
~
L
N
PE
A
Voltmeter
208 V / 230 V
±
10%
MD
1k
Ω
10k
Ω
0.015
μ
F
Voltage source