båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 5
Tab 2
2.1 Help messages
2.1
HINWEIS
i
The above measures clear those help messages that result from operator er-
rors. If it is not possible to clear the help message by taking the above mea-
sures, another type of error is the cause. To identify the error, proceed as de-
scribed in section 2.5
.
H4 07
Unit supports default settings for this exposure,
since the settings specified by SIDEXIS cannot
be taken over.
z
Perform alternative program setting.
H4 10
The sensor is being polled.
z
Wait until the message disappears.
H4 20
The image could not be transferred to SIDEXIS.
VORSICHT
Do not switch the system off until the help mes-
sage has disappeared.
z
Retrieve the exposure with SiRescue (see
SIDEXIS Operator’s Manual).
Help code
Description
Actions required