båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 13
Tab 2
2.5 List of error messages
2.5
Location 07 (XG
Plus
): Easypad/Board DX7
Error code
Description
Actions required
see page
E6
07
01
General error during module initializa-
tion
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
If the error occurs again...
z
Check user interface with electronics (DX7) or
Replace Easypad
E6
07
02
Invalid system data or uninitialized
module storage data
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Check user interface with electronics (DX7) or
Replace Easypad
E6
07
03
Invalid commanding or control data
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
07
04
Data transfer error or dialog error to
module (master side)
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
07
05
Data transfer error or dialog error to
bootloader of module
Note:
Occurs only in connection with
software update
z
Repeat software update
z
Checking the CAN bus
z
Check user interface with electronics (DX7) or
Replace Easypad
E6
07
06
Module failed in TTP*
(detected on master side)
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Replace Easypad
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...