2.5 List of error messages
Tab 2
E1
11
10
The system is operating with default
settings; exposure is possible
Note:
If this error occurs in combination with
errors E1
61
02, E1
91
02 or E1
11
04
(see error logging memory), these
additional errors provide an indication
of the module involved.
z
Read out log memory and take the appropriate
actions
z
Readjust the unit
If the error occurs again...
z
Replace the defective module (DX61, DX91 or
DX11).
E7
11
11
Wrong device configuration
z
Check the device configuration via service rou-
tine S017.2 and reconfigure if necessary
E7
11
12
Internal error in data management of
board DX11
If the error occurs after a module has been
replaced...
z
display "Extended Details" via SIXABCON and
coordinate all further action with the Sirona
Customer Service Center.
If no module has been replaced...
z
switch the unit off, wait for 30 seconds
and then switch it back on.
z
Check software versions via info screen or ser-
vice routine S008.2,
perform a software update if necessary
If the error occurs again...
z
Perform a software update (bug fix)
E6
11
13
Up to HW version AG:
Short circuit in radiation release signal
path between board DX42 and board
DX41 (cable L17).
The release signal was detected on
boards DX11 and DX41, but not on
board DX42.
Up to HW version BA (from 11.2006):
Short circuit in radiation release signal
path between board DX42 and board
DX11 (cables L117 and L108).
The release signal was detected on
board DX11, but not on board DX42
z
see section entitled "Error analysis of X-RAY
control signal path" up to unit hardware version
AG (with board DX41)"
E7
11
14
Wrong remote control for this unit
HINWEIS
i
This error message blocks all unit func-
tions. Work is possible at the unit again
after the remote control is unplugged
and the unit is switched off and on
again.
z
Install the correct remote control.
If necessary, obtain new remote control
from the manufacturer.
(A remote control from GALILEOS or from
an outside manufacturer might have been con-
nected)
Error code
Description
Actions required
see page
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...