båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 21
Tab 2
2.5 List of error messages
2.5
If the error occurs again...
z
Check light barrier V1_1 (X802) with service
routine S015.5,
replace light barrier if necessary
z
Check electric drive and jolt-free running with
service routine S015.5 (free travel),
z
replace the actuators and/or spindles and
replace board DX1 if necessary
E6
13
05
Actuator 1 is not ready for operation
This error is a sequential fault.
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON.
z
Repeat procedure and observe causal error
messages.
E6
13
07
Error when activating actuator 1
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON and check functioning.
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E4
13
11
Actuator 2 has not reached pan home
position
z
Check the actuator mechanisms manually for
smooth and easy running.
In case of stiffness...
z
manually reset the actuators to the zero position
and run a test cycle via P1.
If the error occurs again...
z
Check light barrier V1_2 (X803) with service
routine S015.5,
replace light barrier if necessary
z
Check electric drive and jolt-free running with
service routine S015.5 (free travel),
z
replace the actuators and/or spindles and
replace board DX1 if necessary
E4
13
12
Actuator 2 has not left pan home posi-
tion
z
Check the actuator mechanisms manually for
smooth and easy running.
In case of stiffness...
z
manually reset the actuators to the zero position
and run a test cycle via P1.
If the error occurs again...
z
Check light barrier V1_2 (X803) with service
routine S015.5,
replace light barrier if necessary
z
Check electric drive and jolt-free running with
service routine S015.5 (free travel),
z
replace the actuators and/or spindles and
replace board DX1 if necessary
Error code
Description
Actions required
see page
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...